Surge & Ring wave CDN Features
- Surge testing of unshielded symmetrical telecom lines
- ISO 17025 Calibration & 2 Days Free Transit
- Complies with IEC 61000-4-5 edition 3
- Complete set includes all accessories
- Easy to use bench top housing
- Turnkey Solution - Arrives Ready for Testing!
Overview
Teseq’s CDN 117 coupling-decoupling network is designed for convenient surge testing of telecommunications equipment to IEC/EN 61000-4-5, which specifies a 1.2/50 μs or a 10/700 μs pulse. It includes the special decoupling network and coupling elements that are required for these tests. The design of the CDN is given in IEC 61000-4-5 Ed.3 Figure 10 and A.4. It can be easily interfaced with the EUT and is designed as a bench top unit. It can be used with Teseq’s NSG series or any industry standard surge generator with the appropriate connector adapter.
The user can manually select coupling modes by connecting the generator’s output to the appropriate input. All coupling methods described in the standard can be configured with the CDN 117. Two CDN 117 main units can be arranged in parallel for testing eight line applications (low speed transmission only).
Specifications
Technical specifications |
Teseq CDN 117A-C4-4-1 |
Max operating voltage | AC 250 V, DC 250 V |
Max operating current | 1 A |
Max pulse voltage | CDN 117A-C4-4-1: 4.8kV |
Ohmic resistance per path | <1.0 Ω (typical 0.49 Ω) |
Decoupling chokes 1 kHz: | 20 mH nominal each line |
Transmission 3 dB BW | 1100 Hz (@ 1 kΩ load) |
Connectors | 4 mm safety banana, to generator and EUT |
Pulse |
Surge: 1.2 / 50 μs pulse Ring wave: 0.5 μs / 100 kHz pulse |
Surge coupling |
- 0.5 μF capacitor via 40 Ω - GDT: Gas discharge tube (90V) via 40 Ω - ABG: Avalanche breaking diode (140V) via 40 Ω |
Ring wave coupling | capacitor >=3 μF |
Included Accessories |
|
1 HVS |
HV cable Banana – Banana, 0.5 m, 1 x red, 1 x black |
Earth connection | 1 brass bar 0.3 m including mounting parts, screw M4 x 10 mm |
User manual | Delivered on USB memory card |
Calibration certificate | included |
Protection module options | |
VPM 250-100 |
Varistor Module 250 V, 250 VAC, 350 VDC, Impedance to GND: 100 kΩ |
DPM 50-1 | Module 50 V, bidirectional diodes, Impedance to GND: 1 kΩ per line |
DPM 50 | Module 50 V, bidirectional diodes, Impedance to GND: open circui |