AXOS5 Video Guides
AXOS5 Surge & EFT Testing (includes CCL)
This video includes the generic test setups of the Haefely AXOS5 test system for both combination wave surge testing as well as EFT or burst testing including using the IP4B capacitive coupling clamp (CCL).
Haefely AXOS Platform Overview
This video provided by Haefely, the manufacturer of the AXOS5 and AXOS8 test systems, provides system information and a brief overview of capabilities and connections on front and back.
Data SheetUser ManualOverview
The Haefely Axos 5 EMI (conducted immunity) test generator for 5kV Surge and EFT with a built-in single phase coupling/decoupling network. This compact Haefely product provides safe and reliable operation by using safety interlock, warning lamp and emergency stop functions helping to ensure safe use.
A major advantage of this Immunity Test Generator is the ability to monitor voltage and current through front of the test system. This allows for effective verification of waveform without need for external attenuator. Take advantage of pro-rated daily rates after the first month and quick delivery!
This Immunity Test Generator is designed for EFT / Burst, Surge, Magnetic Field and AC & DC Dips and Interrupts allowing for a diverse range of applications. This pulse generator is easy to operate with manual and automated test mode and software assisted test preparation. The touch screen provides efficient and effective manipulation of testing criteria to meet or a diverse set of testing needs.
The Axos 5 expandable functionality allows for quick and completely automated testing to EN/IEC 61000-4-4, EN/IEC 61000-4-5 and EN/IEC 61000-4-11 and CE Marking. The unit also includes an integrated single-phase coupling decoupling network (CDN) for EMC testing applications up to 16A. The AXOS unit can either be operated via front panel by large color graphic interface or remotely from the PC for added flexibility.
Specifications
Common Standards |
EN/IEC 61000-4-4, EN/IEC 61000-4-5, EN/IEC 61000-4-11 & CE Mark |
General Data |
|
Control Power |
85 V – 264 V, 50/60 Hz |
User test storage |
unlimited |
Remote interface |
Ethernet, RJ45 |
Display |
7“ / 800x480 / 24 bit with touch-screen |
AUX. interface |
8D-sub 37p for external CDN, external transformer etc |
Synch input |
BNC, 10 V – 264 V AC |
External start / stop input |
5 V TTL, starts / stops predefined test sequence |
Dimensions |
45 x 18 x 49cm, 17.11 x 7.02 x 19.29 inches |
Control Power |
85 V – 264 V, 50/60 Hz |
Weight |
25kg, 55 lbs |
IEC 61000-4-4 |
|
Output voltage |
0.2 – 5.0 kV ±10% at coaxial output |
Polarity |
pos / neg / alternate |
Output impedance |
50 Ohms |
Rise time |
5 ns ±30% |
Impulse duration |
50 ns ±30% at 50 Ohm 50 ns –15 +100 ns at 1000 Ohm |
Burst mode |
normal, continuous, real, random |
Spike frequency |
1 Hz – 1 MHz |
Burst duration |
10 us – 1 s |
Integrated single phase coupling / decoupling network |
264 V AC / 16 A 220 V DC / 10 A |
Output voltage |
0.2 – 5.0 kV ±10% at coaxial output |
IEC 61000-4-5 |
|
Output voltage |
0.2 – 5.0 kV ±10% |
Voltage Rise Time / Impulse duration |
1.2 μs ±30% / 50 μs ±20% |
Output Current |
0.1 – 2.5 kA ±10% |
Current Rise Time / Impulse Duration |
8 us ±20% / 20 us ±20% |
Polarity |
pos / neg / alternate |
Output Impedance |
2Ohms & 12Ohms Automatic |
Phase Sync |
0 – 359°with 1°steps or asynchronous mode |
Integrated Single Phase CDN |
264V AC / 16A 220V DC / 10A |
All Accessories
Item | Capabilities |
DIP 116 | Automatic Dips Transformer 16 A 40/70/80% |
FP-EFT 32M |
3-Phase CDN EFT/Burst 32 A / 690 V |
FP-EFT 100M2 |
3-Phase CDN EFT/Burst 100 A / 690 V |
IP4B | Capacitive Coupling Clamp for EFT/Burst |
FP-COMB 32 |
Automatic 3-Phase / 32A CDN Surge and EFT/Burst 2 |
FP-SURGE 100M2 |
3-Phase CDN Surge 100A / 690 V |
PCD 121 |
Symmetrical Data & Control Line Coupler |
PCD 126A |
Asymmetrical Data & Control Line Coupler |
DEC 5 |
Symmetrical Data & Control Line Decoupler |
DEC 6 |
Symmetrical Data & Control Line Decoupler |
DEC 7 |
Asymmetrical Data & Control Line Decoupler |
MSURGE-A |
Magnetic Field Test IEC / EN 61000-4-9 |
VTM 15000 |
Isolation Test 1.2/50 us up to 10 kV |
VTM 15000/05 |
Isolation Test 1.2/50 us up to 10 kV / 0.5J |
PDP 8000 |
HV Differential Probe 1000:1 for Surge |
CP 101 |
Current Probe Model for Surge |
ES |
External Emergency Stop Switch P12 |
WL |
External Warning Lamp P12 |
Common Standards
Axos 5 Testing Standards:
- IEC/EN 61000-4-4 EFT / Burst
- IEC/EN 61000-4-5 Surge Edition 2 & 3
- IEC/EN 61000-4-9 Imp Magnetic Field
- IEC/EN 61000-4-11 AC Dips and Interrupts
- IEC/EN 61000-4-29 DC Dips and Interrupts
- IEC/EN 61000-6-1 Generic Residential
- IEC/EN 61000-6-2 Generic Industrial
- IEC/EN 60335-1 Household
- IEC/EN 60601-1 Medical