Description
Overview
The T3700 Electric-Field Susceptibility Transformer was designed to meet or exceed the voltage requirements of DO 160 Section 19 and Boeing D6-16050 Section 7.2 testing. It may also be used as a 1-to-37 step-up transformer for voltages up to 3700 VRMS.
The T 3700 transformer is able to generate up to a 3700 VRMS potential. It is withstand tested to 135% of rated output to ensure a long, trouble-free life. The turns ratio provides a 1-to-37 step up to the secondary winding.
The T 3700 provides convenient input connectors via touch-proof, four-way binding posts. Standard 0.75-inch spacing of binding posts allows use of standard plugs.
A 1-amp fuse protects the unit against over-current conditions. A rugged, impact-resistant case and robust design protects the transformer from accidental damage.
The T 3700's voltage output meets or exceeds the requirements of DO 160 Section 19 and Boeing D6-16050-5 Section 7.2. It is for use by experienced staff.
Data SheetSpecifications
Common Standards |
DO 160 Section 19 & Boeing D6-16050-5 Section 7.2 standards |
Primary: |
100 VRMS maximum or 1 ARMS, maximum |
Secondary: |
3700 VRMS maximum output |
Physical Turns Ratio: |
1-to-37 step up |
Dielectric Test: |
5 kVDC primary to case 5 kVDC secondary to case |
Frequency Range: |
250 Hz to 35 kHz |
Protection: |
Current-protected input |
Input Terminals: |
4-way binding posts |